Calibration

Calibration

When using a system matrix based reconstruction algorithm one usually uses calibration data to setup the system matrix. In the following we describe measured calibration datasets.

Each calibration dataset is acquired with a small sample of a specific tracer representing an image voxel. The delta sample is placed at various positions within the target FOV and calibration measurements are taken.

Even for 1D and 2D sequences we measured the system matrix on a 3D grid. While this is not strictly necessary for image reconstruction it can be very usefull when studying the particle physics. It is advised to extract appropriate slices prior to reconstruction.

All calibration datasets are also acquired outside the drive-field FOV. This "overscanning" is essential to avoid particles from outside the FOV influencing the reconstruction result inside the FOV. The overscan may be manually shrinked prior to reconstruction

Tracer Perimag (100 mmol), Delta Sample 2mm x 2mm x 1mm Before each measurment a 30 minute Warm Up has been applied

1D Sequence

ParameterValue
Drive-Field Amplitude12 mT x 0mT x 0 mT
Drive-Field Frequency2.5 MHz/102 x 2.5 Mhz/96 x 2.5 Mhz/99
Selection-Field Gradient-1.0 T/m x -1.0 T/m x 2.0 T/m
Sampled Field of View24 mm x 24 mm x 12 mm
Grid size19 x 19 x 19
Repetition Time40.8 $\mu$s
Number of Averages1000
Number of Background Scans362

2D Lissajous Sequence

ParameterValue
Drive-Field Amplitude12 mT x 12 mT x 0 mT
Drive-Field Frequency2.5 MHz/102 x 2.5 Mhz/96 x 2.5 Mhz/99
Selection-Field Gradient-1.0 T/m x -1.0 T/m x 2.0 T/m
Sampled Field of View24 mm x 24 mm x 12 mm
Grid size19 x 19 x 19
Repetition Time652.8 $\mu$s
Number of Averages1000
Number of Background Scans362

3D Lissajous Sequence

ParameterValue
Drive-Field Amplitude12 mT x 12 mT x 12 mT
Drive-Field Frequency2.5 MHz/102 x 2.5 Mhz/96 x 2.5 Mhz/99
Selection-Field Gradient-1.0 T/m x -1.0 T/m x 2.0 T/m
Sampled Field of View24 mm x 24 mm x 12 mm
Grid size19 x 19 x 19
Repetition Time21.54 ms
Number of Averages1000
Number of Background Scans362

Transfer Function

Currently the transfer function of the receive chain is not included in the measured data. However, we plan to include them in the datasets soon.

<!โ€“ All datasets include the transfer function of the receive chain. One has to divide the measured frequency data by this transfer function in order to get corrected signals. This is in particular important when interpreting time signals. โ€“>